Holtgreven GSE-355 I.S. User Manual

Browse online or download User Manual for Scales Holtgreven GSE-355 I.S.. Holtgreven GSE-355 I.S. User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 122
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews

Summary of Contents

Page 1 - GSE Model 350/355

GSE Model 350/355 DIGITAL WEIGH INDICATOR Technical Reference Manual Version 2.0

Page 2

1 CHAPTER 1:INTRODUCTION Thank you for selecting the GSE Model 350/355 Indicator. The Model 350 and Model 355 continue the GSE tradition of Excellence

Page 3 - Model 350/355

91 DISPLAY READS X 100.00 9. Pause for motion delay. DISPLAY READS X Cal ~ Good? ~ 100.00 10. Press( to accept calibration. DISPLAY READS X Ent

Page 4 - GSE Locations

92 8. Enter . DISPLAY READS X 10000 9. Press1 to establish span. DISPLAY READS X 10000. 10. Pause for motion delay. DISPLAY READS X Cal ~

Page 5 - Table of Contents

93 9. Press ( to establish span. DISPLAY READS X 10000. 10. Pause for motion delay. DISPLAY READS X Cal ~ Good? ~ 10000. 11. Press ( to accept calibr

Page 6

94 DISPLAY READS X 2500. 9. Pause for motion delay. DISPLAY READS X Cal ~ Good? ~ 2500. 10. Press 1 to accept calibration. DISPLAY READS X Ente

Page 7

95 11. Press ( to save calibration. DISPLAY READS X Enter ~ =End 12. Press ( to exit calibration. DISPLAY READS X 5055. 13. Remove the calibration

Page 8

96 DISPLAY READS X Fast ~ Cal~First ~ Zero? ~2640. 3. Press ! ! !. DISPLAY READS X Only ~ Zero? ~ 2640. 4. Remove any load on the scale. DISPLAY READ

Page 9

97 5. Place a 100lb test weight on scale. DISPLAY READS X Code03 6. Press . DISPLAY READS X First ~ Zero? ~ -0.26 7. Remove any load on the sca

Page 10 - CHAPTER 1:INTRODUCTION

98 4. Press ( to establish zero. DISPLAY READS X Enter ~ Load? ~ 0.00 5. Place a 100lb test weight on scale. DISPLAY READS X Code03 6. Press ".

Page 11

99 If the calibration weight value was entered before applying the test weight, Add ~ Load is displayed indicating that the test weight should now be

Page 12 - Display

100 3. Press 1 or (. DISPLAY READS X Enter ~ =End 4. Press 1 or ( to exit calibration. DISPLAY READS X 99.66  When saving calibration, param

Page 13 - 350 KEYPAD

2 Input (J4): 12 – 26VAC / 12 – 36 VDC One of four wall mount transformers supplied: North American: Input: 120VAC, 30W, 60 Hz Output: 20VAC, 800mA Un

Page 14 - 355 Keypad

101 CHAPTER 6: LEGAL FOR TRADE The Model 350/355 default parameter setup does not ensure compliance with legal-for-trade installations as mandated by

Page 15 - Weigh Mode Functions

102 To install the panel mount tag: 1. Remove the electronics. 2. Carefully score the case so the drill will not slip. 3. Drill the mounting holes

Page 16

103 Sealing And Audit Trails Most legal-for-trade installations will require the Model 350/355 to be sealed. A sealed indicator cannot be accessed for

Page 17 - Mounting

104 AUDIT TRAILS Three separate incrementing, non-resetable audit trail parameters are used by the Model 350/355 to indicate changes to various parame

Page 18 - OUTLINE DRAWINGS

105 DISPLAY READS XSetup~Enter ~ Code! 2. Press ( DISPLAY READS X -No- ~ Chgs!~P110 ~ FS ~ xx.xx 3. Press  !. DISPLAY READS X Audit ~ Trail~C

Page 19

106 CHAPTER 7: TROUBLESHOOTING This chapter contains error messages and information parameters, as well as information on setup parameter selections a

Page 20 - SERIAL PORT CONNECTIONS

107 SETUP MODE ERRORS Bad ~ Code! An incorrect access code was entered. Unit ~ Seald Access to the Setup or Calibration Mode was denied. Check

Page 21 - (Die cast encl. only)

108 CALIBRATION ERRORS F.S. ~ TooHi The entered calibration weight will result in an over-capacity condition at full scale. Verify that the full sc

Page 22 - POWER CONNECTION

109 To view the setup parameter selections (355): 1. From the Weigh Mode, press  !. DISPLAY READS X Setup~Enter ~ Code! 2. Press (. DISPLAY READS

Page 23 - 12-36 VDC

110 Table 29: Information Parameters PARAMETER NAME DESCRIPTION 60000 E2Ins Total amount of EEPROM storage. 60001 E2Fre Amount of available EEPROM sto

Page 24

3 Wall Mount Kit Swivel bracket/stand for zinc die cast enclosure Panel Mount Kit Mounts zinc die cast enclosure to user panel Cutou

Page 25 - Panel Mount Kit

111 62005 Analg ~ 0-20A 62006 Analg ~ 4-20A Allows the analog output to be changed by pressing 1 or (. Output will toggle through 0, 25, 50 and 100 p

Page 27 - Analog Card Connections

Part Number 39-10-41432

Page 28

4 ANNUNCIATORS Annunciators provide mode and status information. When illuminated, they indicate the following conditions: SP1 Setpoint #1 activa

Page 29

5 Key Press WEIGH MODE Count Mode Setup Mode 1 Performs an auto-tare function (if enabled) and/or accepts an entry in progress. Performs an auto-t

Page 30

6 Key Press WEIGH MODE Count Mode Setup Mode " Performs a gross zero function and/or clears an entry in progress. Hold this key on power-up

Page 31 - 0-20A ~ 50P

7 2 Toggles the units of measure between lb, kg, g, lb oz, oz. # Tares any displayed weight and selects the net mode. ! Toggles the display between GR

Page 32 - Analog Output Setup

8 7.33” – 7.45”(186.2 – 189.2mm)5.25” – 5.37”(133.4 – 136.4mm)PANELCUTOUTCHAPTER 2: INDICATOR INSTALLATION This chapter contains information necessary

Page 33 - Analog Output Calibration

9 PERMANENT MOUNTING The optional mounting bracket allows the zinc die cast enclosure to be securely fastened to another surface. The bracket is attac

Page 34 - 1 (350) or ( (355) key

10 Figure 7: Model 350 Stainless Steel Dimensions Figure 8: Model 355 Outline Drawing

Page 36

11 Wiring A description of all wiring terminals is included on the bottom label of the zinc die cast enclosure as shown in Figure 9. LOAD C

Page 37

12 The maximum recommended cable length is 50 feet (15 meters). However, much longer connections are possible if using a properly shielded, low-capaci

Page 38 -  !

13 well. A minimum contact duration of 100 ms is required. Once invoked, the selected remote key operation will not repeat until the contact is releas

Page 39 - RS-485 Networking

14 INTERNAL AC OR DC POWER SUPPLY The Model 350/355 internal power supply model may be powered by an external VAC power source connected to (J10) on t

Page 40 - NETWORK CONNECTIONS

15 CHAPTER 3: OPTION INSTALLATION The capabilities of the Model 350 can be expanded with the use of one or more option kits. This chapter provides in

Page 41

16 Swivel Bracket The Model 350/355 has an optional stainless steel swivel bracket for secure mounting to desks, tabletops or walls. See the Mounting

Page 42 - 20 mA Current Loop Option

17 Figure 18: Model 350 Panel Mount Installation (Die Cast Model) o INDICATOR CAN BE INSTALLED USING EITHER BUTTON OR HEX HEAD SCREWS.

Page 43 - CONNECTED DEVICES

18 Analog Card Connections The 16-bit electrically-isolated analog output module provides a highly accurate analog signal, proportional to weight. Thi

Page 44 - Battery Power Supply

19 To install the Analog Output Model (350 Stainless Model and 355): 1. Open the indicator. Remove the eight screws from the back of the unit. 2.

Page 45 - PIN DESIGNATIONS

20 Table 2: Analog Output Connections Pin Connection Name Description 1 Earth Ground Non-isolated earth ground (future use). 2 + 5 VDC Isolated 5 v

Page 46 - BATTERY CHARGING

Model 350/355 Technical Reference Manual Version 2.0

Page 47 - AUTOMATIC SHUTDOWN

21 To test the 0-10v output mode: 1. Enter the Setup Mode (see Setup Mode on page 40). Chngs Poss! P110.-- ~ F.S.= ~ 100.00 2. Attach the voltmet

Page 48 - Splash Guard Protection

22 6. Press 1 or ( to set the output to 0%. 0-20A ~ 0P 0.00 V 7. Press 1 or ( to increase the output to 25%. 0-20A ~ 25P 2.5 V 8. Press1 or ( to i

Page 49 - CHAPTER 4: CONFIGURATION

23 6 V 9. Press 1 or( to increase the output to 75%. 4-20A ~ 75P 8 V 10. Press 1 or ( to increase the output to 100%. 4-20A ~ 100P 10V Analog Ou

Page 50

24 Analog Output Calibration Table 4: Analog Output Calibration Parameters PARAMETER SETTING DISPLAYED NAME DESCRIPTION TYPE/CHOICES (*=DEFAULT) P

Page 51

25 61203%s54555%e P61203. 0 Gn 54555 61204%s10910%e P61204. 4 off 10910 61205%s54555%e P61205. 4 Gn 54555 612

Page 52 - SELECTION PARAMETERS

26 Setpoint Card Connections Using one of the software setpoint configurations (see General Setpoint Setup on page 63) in conjunction with the setpoi

Page 53

27 Be sure to avoid ‘pinching’ the cable between the housing halves. Affix all appropriate labels to the back of the indicator. To install the Setp

Page 54 - to complete exit

28 Table 5: Setpoint Output Specifications (24350B-100C0) Characteristic Limits Operating Voltage Range (47-63 Hz) 20-280 VRMS Max Load Current 1 A

Page 55

29 Place this label above the setpoint card knockout (die cast) Figure 23: Wire the option connector in accordance with the label (left to right)

Page 56 - List of Parameters

30 RS-485 Networking The Model 350/355 controller supports address recognition for multi-drop communications. This section describes the installation

Page 57 - PARAMETER MAP DETAILS

GSE 350/355 Digital Weigh Indicator Technical Reference Manual Copyright © 2004 GSE Scale Systems. All rights reserved. Published by: GSE Scale Sy

Page 58

31 NETWORK CONNECTIONS Apply the supplied label over the appropriate COMM PORT pin designations. On the stainless steel enclosure, apply the label m

Page 60 - " key upon power up

33 20 mA Current Loop Option Description: This option will convert the comm port of the indicator to a 20 mA current loop instead of an RS-232. This

Page 61 - & key

34 MAX VOLTAGE Active mode Tx current loop provides a driving voltage of 12v. This will allow 20 mA current flow with up to a 600 ohm load. Passive

Page 62 - Preset Transmit Selections

35 Battery Power Supply The Battery Module gives the Model 350/355 portability, providing an Model 350/355 with an LCD display and one loadcell 20 hou

Page 63

36 Figure 27: Battery Option Installation (350 die cast) PIN DESIGNATIONS See Figure 28 for main board battery connectio

Page 64 - Custom Transmit

37 Figure 28: Model 350/355 Main Board Battery Connection ON/OFF SWITCH The on/off switch for the Model 350 die cast is a rocker-type switch that mo

Page 65 - ENTERING ASCII CONTROL CODES

38 TEMPERATURE SENSING If the ambient temperature sensor RT1 on the board is colder than 10 degrees C, the battery will trickle charge and not enter f

Page 66 - HEX

39 Version 1 kits (GSE P/N: 24350B-120B0-- Zinc Die Cast Model and GSE P/N: 24350B-121B0-- Stainless Steel Model) do not support automatic shutdown.

Page 67 - %e 0 %e%e 0 %e%e

40 CHAPTER 4: CONFIGURATION Entering the Setup Mode (Model 350) To prevent accidental changes to the Indicator Setup, a sequence of keystrokes is us

Page 68 - %z%c%e%e

i Table of Contents CHAPTER 1:INTRODUCTION... 1 STAN

Page 69

41 Entering the Setup Mode (Model 355) To prevent accidental changes to the Indicator Setup, a sequence of keystrokes is used to gain access to the

Page 70

42 DISPLAY READS X. 2. Press  DISPLAY READS X P111.09 ~ 1Grad ~ 0.01 3. Repeat  to back up one parameter. To access the previous parameter (M

Page 71 - ( key will cause the

43 Changing a Parameter Value SELECTION PARAMETERS Selection parameters have a pre-defined list of choices to pick from. Each choice is numbered and c

Page 72 - Setpoint Setup

44 To enter a Key-In Parameter (355): 1. Key in the number by using the numeric keys. 2. Press ! to accept the entry. To setup a full scale value of

Page 73 - # will abort the

45 3. Press 1 to save setup changes. DISPLAY READS X Enter ~ =End 4. Press 1 to complete exit. DISPLAY READS X 0.00 To exit the Setup Mode and s

Page 74 - EXAMPLE (GENERAL)

46 To exit the Setup Mode without saving changes (355): 1. Press " to begin exiting Setup Mode. DISPLAY READS X Enter ~ =CAL! 2. Press  to byp

Page 75 - BARGRAPH (GENERAL)

47 List of Parameters The Model 350 and Model 355 have several parameters that can be configured to your specific application. Below is a table of the

Page 76

48 Parameter Number Display Name Default Value Valid Range/ Choices Parameter Description Page P410.-- Euro Disable Enable / Disable 9991/9990 (Key

Page 77 - • Target ]

49 Stability is defined as weight fluctuations within an aperture that can be regarded as being a stable weight. Deviations outside this aperture are

Page 78 - EXAMPLE (FILL)

50  Note that if a setpoint activation method is set to 1, disabling Auto Tare will also disable the activation of that setpoint. P169 Auto Tare Cl

Page 79 - PAUSE FEATURE (BATCH)

ii Isolation ... 33 M

Page 80 - Discharge

51 Enable / disable the RS-485 multi-drop network option. Requires that an RS-485 option board be installed. This option allows up to 250 RS-485 devic

Page 81 - PAUSE FEATURE (DISCHARGE)

52 Determines the TIME format style, 12 hour or 24 hour. If in 12 hour mode the right most decimal point on the LED display will become the PM indicat

Page 82 - EXAMPLE (DISCHARGE)

53 Enables or disables the bargraph display on the LCD version of the Model 350/355 display. This setup parameter only appears if an LCD display is i

Page 83 - ACTIVATION METHOD (BOTH)

54 Choice 7: <STX><Unsigned DATA><sp><CR> Choice 8: <STX><Signed DATA><sp><lb/kg><sp><Gr

Page 84 - EXAMPLE (BOTH)

55 Custom Transmit The serial output of the Model 350/355 can be configured for a custom application such as a remote display format, a customized com

Page 85

56 Figure 29: Custom Transmit File ENTERING ASCII TEXT ASCII text is defined as printable characters, including alpha-numerics as well as punctuation

Page 86 - Bargraph Setup (LCD)

57 Table 12: ASCII / HEXADECIMAL CONVERSION CHART HEX CHAR DEC 00 NUL 000 1A SUB026 34 4 052 N 078 68 h 104 0

Page 87

58 PARAMETER SELECTION NUMBERS The following sequence enters parameters into a custom transmit: %e , the parameter number, %e%e , a format code, and t

Page 88

59 Time/Date Format Codes 1 “10:07:40 am 01/26/01 Includes seconds with time. 2 “10:08 01/26/01” 24 hour time format. 4 “10:11 am Jan 26, 2001” Print

Page 89 - CHECK-WEIGHING)

60 DISPLAY READS X 20 To sample using variable counts: 1. From the Weigh Mode, Press ! to view the current net weight. DISPLAY READS X 1.05 2.

Page 90 - Description

iii Pre-acts (General) ... 64 Learn F

Page 91 - EIGHING)

61 Analog Output Calibration  This section requires firmware rev 010 or later. Table 16: Analog Output Calibration Parameters PARAMETER SETTING DIS

Page 92 - Remote Serial Operation

62 61202%s0%c P61202. 0 off 0 61203%s54555%e P61203. 0 Gn 54555 61204%s10910%e P61204. 4 off 10910 612

Page 93 - DISPLAY CAPTURE UTILITY

63 Setpoint Setup The Model 350/355 has several pre-programmed scale setpoint applications available at P5100. Various related setpoint parameters may

Page 94

64 PRE-ACTS (GENERAL) Pre-Acts are control actions prior to reaching a desired target value. Necessary to prevent over-filling due to product flow rat

Page 95

65 CHANGING TARGETS FROM THE WEIGH MODE (GENERAL) When a setpoint operation is configured from the Setup Mode, certain parameters are made available i

Page 96 - NETWORK PROTOCOL

66 To change targets from the Weigh Mode (355 IS): 1. Press $. DISPLAY READS X0.00 2. Press ! to view the current fill target. DISPLAY READS XTarg

Page 97 - CHAPTER 5: CALIBRATION

67 NOTE: Right-side bars will have the same scaling (lb per bar) as the left-side bars. They are represented as a mirror image of each other. Bars O

Page 98 - FIRST ZERO

68 Fill The fill program is used for single-speed or dual-speed filling operations. The dual-speed fill operation allows for both a fast and a slow fi

Page 99

69 Pre-act 2 specifies the target where the final cutoff should occur, regardless of a single-speed or dual-speed operation.  Pre-act 1 controls set

Page 100 - LAST ZERO

70 Setpoint Selection Function Description P5107.0 Start 1 Setpoint 1 activation method. P5108.-- Targ 2 Ingredient 2 target value. P5109.-- PA 2

Page 101

iv Example (Absolute Check-Weighing) ... 83 REMOTE KEY OPERATION..

Page 102 - FALSE ZERO

71 PA 2 are the subsets for Targ 1. The pre-acts can be deleted as subsets by choosing ‘disabled’ at P5114 and P5115. See Changing Targets from the We

Page 103

72 Table 20: Setpoint Setup (Discharge) Setpoint Selection Function Description P5100.1 Discharge Select discharge setpoint operation. P5101.-- Tar

Page 104 - ONLY ZERO

73 TARGET CHANGES FROM THE WEIGH MODE (DISCHARGE) When Discharge is configured in the setpoint setup, Targ 1 automatically becomes an available mode f

Page 105 - RESET CALIBRATION

74 Setpoint Selection Function Description P5105.-- PA 1 Pre-act 1 value for vessel fill. P5106.1 Learn 1 Learn feature for pre-act 1. P5107.0 S

Page 106

75  Pre-acts 1 and 2 are always available as subsets of their respective targets from the Weigh Mode. Since the discharge portion of the Both cycle

Page 107

76 Setpoint Selection Function Description P5137.1 Rstb 2 Reset 2 stability setting. P5141.1 Base 3 Select from Net, Gross or Count (Quantity) fo

Page 108

77 EXAMPLE (INDEPENDENT) Setup a continuous-cycle static weighing system that fills a weigh hopper from a storage bin. The weigh hopper should stop th

Page 109 - 1 or ( to exit calibration

78 8 40.00 to 41.25 7 41.25 to 42.50 6 42.50 to 43.75 5 43.75 to 45.00 4 45.00 to 46.25 3 46.25 to

Page 110 - CHAPTER 6: LEGAL FOR TRADE

79 Figure 35: Examples of the Bargraph "Below, Within and Above Tolerance" NOTE: (bargraph for filling and emptying modes) For modes

Page 111 - Other Requirements

80 according to the percentages entered. The desired target may be based on gross weight, net weight or quantity (if counting is enabled). Only the an

Page 113 - AUDIT TRAILS

81 3. Press  1 or  ( to set a new fill target of 400 lbs. DISPLAY READS XTarg1 ~ 400 4. Press 1 or ( to select the Pre-Act 1 (subset) value of

Page 114 -  !

82 SETPOINT ACTIVATION (ABSOLUTE CHECK-WEIGHING) In order for the annunciators or setpoints to activate, the displayed value must be at least five gra

Page 115 - CHAPTER 7: TROUBLESHOOTING

83 EXAMPLE (ABSOLUTE CHECK-WEIGHING) With a system set up to check-weigh 50 lb. cement bags, the following settings might be used check-weigh bag from

Page 116 - HARDWARE ERRORS

84 Command ASCII HEX Description Piece Wt. Entry %9 B9h Allows a piece weight to be entered serially. For example, 0.10%9 will enter a piece weig

Page 117 - Viewing Setup

85 2. Press 2 twice to move the decimal point over to separate the month from the day. It is not necessary to enter a 0 before a single digit month. I

Page 118

86 7. Press  to toggle through the numbers to enter the seconds. The seconds do not have to be entered. Press 1 to bypass entering the seconds. DIS

Page 119 - 1 or ( while

87 The RS-485 multi-drop network option supports both half duplex (2- wire) and full duplex (4-wire) modes of operation. The mode of operation is det

Page 120 - A/D Calibration Procedure

88 CHAPTER 5: CALIBRATION Setup Mode Calibration You can enter the Calibration Mode after accessing the Setup Mode to view and/or change parameter set

Page 121

89 Establishing Zero The Model 350/355 provides five methods for obtaining a zero (no load) calibration reference, First Zero, Last Zero, False Zero,

Page 122

90 DISPLAY READS X Enter ~ Load ~ 0.00 6. Place a 100lb test weight on scale. DISPLAY READS X Enter ~ Load ~ 99.66 7. Enter . DISPLAY READS

Comments to this Manuals

No comments